Strain-Gauge Load Cell: Calibration & Uncertainty
Built and calibrated a strain-gauge cantilever scale, then compared beam-theory predictions with measured loads and tested how consistently it weighed a water bottle.
Mechanics of materials laboratory · Calibration, instrumentation, and uncertainty
My contribution
I bonded and wired the gauge, balanced the quarter bridge, and used LabVIEW to convert amplified voltage into strain and weight. I characterized four reference loads, repeated ten bottle placements, and analyzed calibration and measurement uncertainty using least squares, RSS propagation, and Monte Carlo simulation.
Results & observations
The four-load fit gave a calibration slope of −9.76 ± 0.45 N/V. Ten bottle placements averaged 2.50 ± 0.16 N by calibration, compared with a 2.4759 N balance reference; placement standard deviation was 0.017 N. The uncertainty intervals use the course’s approximate 95% coverage assumptions.
Building and calibrating the load cell
I built a scale around an aluminum cantilever and a bonded strain gauge and compared two ways of finding weight: a mechanics model based on beam geometry and a calibration fitted to four labeled reference loads.
Getting the signal into range
The quarter bridge and amplifier turned a small resistance change into a voltage LabVIEW could read. I checked gauge resistance, soldered the leads, balanced the bridge, centered the amplifier, and tared the unloaded beam. Acquisition used 2,500-sample blocks at 50,000 samples per second, with voltage windows selected to avoid saturation while retaining resolution.
Calibration and Monte Carlo uncertainty
The reference loads spanned approximately 1.05–3.93 N, giving a fit with R² = 0.9998. I used 120,000 Monte Carlo trials to propagate coordinate uncertainty into the calibration slope. The saved calibrated-weight columns were zero, so I calculated those weights afterward from the recorded voltage and fitted slope because they were not captured as live instrument output.
Repeatability and bottle-weight measurements
Ten bottle placements gave a 0.017 N sample standard deviation, but common tare and slope uncertainty did not disappear through averaging. The calibrated mean was 2.50 ± 0.16 N and the mechanics mean was 2.50 ± 0.11 N, both consistent with the balance reference. Four measured gulps ranged from 0.45 to 0.81 N; the empty-bottle signal was smaller than its uncertainty.
Measurement range and sensitivity
The report estimated an amplifier-window limit near 6.3 N and first yield near 42 N. Both are extrapolations, not tested capacities. Original tare statistics were unavailable, so the analysis used a conservative proxy from unloaded measurements. A sensitivity check also showed that using the package gauge factor instead of the course setting would raise mechanics-based weights by 4.25%.